  Phoenix AM pacemaker is 8 Parameter multi-programmable Implantable Pulse Generator. Phoenix AM has Digital Algorithm for automatic threshold measurement facility for tension free pacing even if the threshold is changing between acute and chronic levels. Once automatic feature is enabled the pacemaker itself settles down to an optimum pacing amplitude. When a loss of capture is detected pacemaker immediately increases its amplitude up to 4.5 Volts to maintain pacing, after 4.5Volts Automatic feature is disabled and pacing polarity is changed to Unipolar for effective pacing (eliminating defect caused by outer lead fracture).
   
 
Specifications
   
  Programmable Parameter Values
   
 

Pacing Modes

VVI, VVT, VOO, OVO, AAI, AAT, AOO, OAO

Auto Measure Switch

ON / OFF

Basic Pacing Rates

From 30 to 120 ppm in steps of 2 ppm

Pulse Widths

21 values from 0.06 ms to 1.50 ms

Pulse Amplitudes

32 values from 0.2 to 7.5 V

Refractory Periods

From 203 to 438 ms in steps of 15.625 ms

Sensitivities

From 0.5 to 4.0 mV in steps of 0.5 mV

Hysteresis

From 0 to 20% in steps of 5%

Pacing Polarities

Unipolar / Bipolar

Sensing Polarities

Unipolar / Bipolar

Upper rate in trigger mode

From 80 to 183 ppm

  Tachycardia Treatment
 

Temporary programming

Asynchronous pacing with rate from 30 to 427 ppm
  Magnetic Response
 

Mode

Voo

Rate

BOL (begin of life) :95ppm
ERI (elctive replacement): 85 ppm

   
 

Noise detection: When the pacemaker senses events with frequency exceeding 11 Hz, it switches to VOO mode.

   
 

End of Service Indicators

 

ERI (Elective replacement): When the remaining capacity of the battery is between 4 and 8%, the pacing rate switches to 85 ppm when a magnet is applied.

EOL (End of Life): When the remaining capacity of the battery is under 4% of the initial capacity, the pacing rate slows 10 ppm.

  Power Source
  Battery Chemistry: Lithium – Iodine WG 9085
Initial Voltage: 2.8 V.
Maximum Available Capacity: 0.98 A-hr.
  Physical Characteristics
  Dimensions:  49 x 38 x 7 mm
Mass: 24 g.
Case Material: Titanium
Connectors: 3.2mm, IS-1
  Service Life
  The expected life of the generator, pacing 100% at 70 ppm, with 2.5 V amplitude and pulse width of 0.488 ms, is 8.2 years.
  Telemetry Data
 

Parameters: All the programmable parameters.
Identification:
Model and serial number
Battery condition: Voltage, ERI and EOL indicators
Statistics: Keeps track of pacing and sensing events, as well as ventricular premature events. Frequency Histogram.

  • Lead Impedance
  • Marker Pulses
  • Temporary Programming
  Protections
  Noise detection: When the pacemaker senses events with frequency exceeding 11 Hz, it switches to VOO mode.
Run Away Limit: The pacemaker includes an antirunaway circuit to avoid the generation of stimuli with a rate over 200 ppm (except in temporary programming).
  Programming
 

The pacemaker is programmed using a microcomputer (usually a Laptop) that runs Windows operating system. It is required to use ALUS programming interface and Wand for programming custom made for Biopace Pacemakers. It also permits the measurement of the stimulation and sensing thresholds as well as the generation of printed reports.